MODEL | DCN1000 W2/L2 | DCN1000 W4/L4 | DCN1000 W7/L7 | DCN1000 W12/L12 |
Emissive area | 50 mm x 50 mm | 100 mm x 100 mm | 180 mm x 180 mm | 300 mm x 300 mm |
Temperature range (L type) | ||||
• Absolute (for any ambient T) | -40°C to +150°C | |||
• Differential (20°C ambient) | -60°C to +130°C | |||
Temperature range (W type) | ||||
•
Absolute (for a 20°C ± 2°C wate circulation) |
-20°C to +150°C | -10°C to +150°C | -5°C to +150°C | |
• Differential (20°C ambient) | -40°C to +130°C | -30°C to +130°C | -25°C to +130°C | |
Thermal uniformity at ambient ±5 °C | 0.01°C | |||
Thermal uniformity at 50°C | 0.1°C | 0.2°C (W) / 0.15°C (L) | ||
Emissivity/ | 0.98 ±0.02 (Option: 0.99 ±0.01) | |||
Apparent emissivity after calibration | 1 | |||
Stability | 0.5 mK over [0;65°C] , 2mK outside | |||
Temperature measurement accuracy | Differential mode : 0.01°C / absolute mode : 0.03°C | |||
Display resolution | 0.0001°C (actual temperature and set point display) | |||
Slew rate (L type only) | > 0.4°C/s heating ; > 0.2°C/s cooling | |||
Stabilization time | 30 seconds | |||
Operating temperature | Control unit: +5°C to + 45°C ; Head: -20°C to +70°C | |||
Remote control | Ethernet, RS232 and IEEE488 interface | |||
Max. power consumption | ||||
• Control Unit | 800 W | 1600 W | 1600 W | |
• Refrigerated Unit (L type only) | 2300 W | 2500 W | 5000 W | |
Head dimensions W x H x D | 115x200x111 mm3 | 192x210x120 mm3 | 247x410x112 mm3 | 370x531x112 mm3 |
Head weight | 2 kg | 5 kg | 10 kg | 20 kg |
Electronic unit size | 2U x 19” | |||
Electronic unit weight | 6.5 kg | 8.5 kg | ||
Double head option | NO | YES | NO | NO |