  • 기술사양
    SPECIFICATIONS: ClareHAL 104, 103, 102, 101, 100


    Power Requirement

    AC Voltage : Selectable 115 or 230 VAC RMS
    Frequency : 50/60 Hz


    Mechanical Specification

    Size (HxWxL) : 300mm-200mm-370mm
    Weight : 15kg Approx


    Ground Bond Test - 104, 103, 100

    Test Voltage : Nominal 6V AC
    Frequency : 50 or 60 Hz (Independent of supply)
    Display Range and Accuracy : 0-1500mΩ ±2% ±5 Counts
    Display Resolution : 1 mΩ
    Compliance Test Ranges : Current/Load Resistance/Time 5 A-1000mΩ - Continuous
    10 A-500mΩ - Continuous
    25 A-200 mΩ - Continuous
    30 A-150 mΩ - 60 secs
    40A-100 mΩ - 60 secs
    Settable Output Current Range : 0.1A-40.0A
    Selectable Range of Pass/Fail Levels : 0-1500mΩ


    Insulation Resistance Test - 104, 103, 102, 101

    DC Output Voltage : 250V, 500V, or 1000V Selectable
    Display Range : 0.01MΩ-500MΩ
    Display Range / Accuracy : 0.03MΩ-350MΩ ± 5% ± 5 Counts


    350MΩ-500MΩ Indication Only

    Display Resolution : 0.01MΩ
    Pass/Fail Level : 0.00MΩ-500.0MΩ


    AC Flash / Hipot Test - 104, 103, 102, 101

    Programmable Voltage Range : 0.10kV-5.00kV (10V/Step Resolution)
    Frequency Independent of Supply : 50 or 60 Hz
    Voltage Display Range and Accuracy : 0.10kV-5.00kV ±1% ± 5 Counts
    Voltage Display Resolution : 0.01kV
    Current Display Range and Accuracy : 0.01mA-20.00mA ±1% ± 5 Counts
    Current Display Resolution : 0.01mA
    Selectable Range of Pass/Fail Levels : 0.01mA-20.00mA
    Maximum Current Output : 20.00mA @5kV
    Optional Arc Detection : 9 Levels


    DC Flash / Hipot Test - 104, 103, 102, 101

    Programmable Voltage Range : 0.10kV-6.00kV (10V/Step Resolution)
    Voltage Display Range and Accuracy : 0.10 kV-6.00kV ±1% ± 5 Counts
    Voltage Display Resolution : 0.01kV
    Current Display Range and Accuracy : 0.01mA-10.00mA ±1% ± 5 Counts
    Current Display Resolution : 0.01mA
    Selectable Range of Pass/Fail Levels : 0.01mA-10.00mA
    Maximum Current Output : 10.00mA
    Optional Arc Detection : 9 Levels


    Power and Leakage Test - 104 only


    Power Output Rating

    Test Voltage : 110V - 230V AC
    20A nominal (Vin-Vout)
    Maximum Power Output : Up to 5.0kVA (dependant on mains supply rating)


    Output Power Measurement


    Single Phase Power Measurement

    Display Range and Accuracy : 0.02kVA-5.00kVA ±2% ±0.02kVA
    Selectable Range of Pass/Fail Levels : 0.001kVA-5.000kVA


    Single Phase Power Factor Measurement

    Display Range and Accuracy : Ratio 0.000 -1.000 ± 0.030
    Selectable Range of Pass/Fail Levels : 0.000-1.000


    Leakage Current Measurement

    Display Range and Accuracy : 0.10mA-20.00 mA ± 1% ±5 Counts
    Display Resolution : 0.01mA
    Selectable Range of Pass/Fail Levels : 0.01mA-20mA


    Touch Current Measurement

    Display Range and Accuracy : 0.02mA - 5.00 mA ± 1% ±5 Counts
    Display Resolution : 0.01mA
    Selectable Range of Pass/Fail Levels : 0.02mA-5.00mA
    Measuring Device : IEC/EN 60990 Fig 4

    (IEC/EN 60990 Fig 3 or Fig 5 Upon Request)