SPECIFICATIONS: ClareHAL 104, 103, 102, 101, 100
Power Requirement
AC Voltage : Selectable 115 or 230 VAC RMS
Frequency : 50/60 Hz
Mechanical Specification
Size (HxWxL) : 300mm-200mm-370mm
Weight : 15kg Approx
Ground Bond Test - 104, 103, 100
Test Voltage : Nominal 6V AC
Frequency : 50 or 60 Hz (Independent of supply)
Display Range and Accuracy : 0-1500mΩ ±2% ±5 Counts
Display Resolution : 1 mΩ
Compliance Test Ranges : Current/Load Resistance/Time 5 A-1000mΩ - Continuous
10 A-500mΩ - Continuous
25 A-200 mΩ - Continuous
30 A-150 mΩ - 60 secs
40A-100 mΩ - 60 secs
Settable Output Current Range : 0.1A-40.0A
Selectable Range of Pass/Fail Levels : 0-1500mΩ
Insulation Resistance Test - 104, 103, 102, 101
DC Output Voltage : 250V, 500V, or 1000V Selectable
Display Range : 0.01MΩ-500MΩ
Display Range / Accuracy : 0.03MΩ-350MΩ ± 5% ± 5 Counts
350MΩ-500MΩ Indication Only
Display Resolution : 0.01MΩ
Pass/Fail Level : 0.00MΩ-500.0MΩ
AC Flash / Hipot Test - 104, 103, 102, 101
Programmable Voltage Range : 0.10kV-5.00kV (10V/Step Resolution)
Frequency Independent of Supply : 50 or 60 Hz
Voltage Display Range and Accuracy : 0.10kV-5.00kV ±1% ± 5 Counts
Voltage Display Resolution : 0.01kV
Current Display Range and Accuracy : 0.01mA-20.00mA ±1% ± 5 Counts
Current Display Resolution : 0.01mA
Selectable Range of Pass/Fail Levels : 0.01mA-20.00mA
Maximum Current Output : 20.00mA @5kV
Optional Arc Detection : 9 Levels
DC Flash / Hipot Test - 104, 103, 102, 101
Programmable Voltage Range : 0.10kV-6.00kV (10V/Step Resolution)
Voltage Display Range and Accuracy : 0.10 kV-6.00kV ±1% ± 5 Counts
Voltage Display Resolution : 0.01kV
Current Display Range and Accuracy : 0.01mA-10.00mA ±1% ± 5 Counts
Current Display Resolution : 0.01mA
Selectable Range of Pass/Fail Levels : 0.01mA-10.00mA
Maximum Current Output : 10.00mA
Optional Arc Detection : 9 Levels
Power and Leakage Test - 104 only
Power Output Rating
Test Voltage : 110V - 230V AC
20A nominal (Vin-Vout)
Maximum Power Output : Up to 5.0kVA (dependant on mains supply rating)
Output Power Measurement
Single Phase Power Measurement
Display Range and Accuracy : 0.02kVA-5.00kVA ±2% ±0.02kVA
Selectable Range of Pass/Fail Levels : 0.001kVA-5.000kVA
Single Phase Power Factor Measurement
Display Range and Accuracy : Ratio 0.000 -1.000 ± 0.030
Selectable Range of Pass/Fail Levels : 0.000-1.000
Leakage Current Measurement
Display Range and Accuracy : 0.10mA-20.00 mA ± 1% ±5 Counts
Display Resolution : 0.01mA
Selectable Range of Pass/Fail Levels : 0.01mA-20mA
Touch Current Measurement
Display Range and Accuracy : 0.02mA - 5.00 mA ± 1% ±5 Counts
Display Resolution : 0.01mA
Selectable Range of Pass/Fail Levels : 0.02mA-5.00mA
Measuring Device : IEC/EN 60990 Fig 4
(IEC/EN 60990 Fig 3 or Fig 5 Upon Request)