MODEL | ECN100 H6 | ECN100 H12 | ECN100 N20 | ECN100 N40 |
Emissive area | 150 x 150 mm | 300 x 300 mm | 500 x 500 mm | 1000 x 1000 mm |
Temperature range | 50°C to 600°C | 50°C to 550°C | 50°C to 300°C | Ambiant +5°C to 100°C |
Emissive area uniformity | 0.75 °C at 300 °C | 2 °C at 300 °C | 3.5 °C at 300 °C | 1°C at 100 °C |
Emissivity / Apparent emissivity after calibration |
0.98
±0.02 over 8 - 14 μm / 1.00 0.96 ±0.02 over 3 - 5 μm / 1.00 |
0.98 ±0.02 / 1.00 | ||
Stability | 0.02°C | 0.05°C | 0.01°C | |
Temperature measurement accuracy |
± 0.5°C | ± 2.5°C | ||
Display resolution | 0.01°C | 0.1°C | ||
Warm-up time from ambient to Tmax |
45 min | 60 min | 30 min | |
Head dimensions W x H x D (mm3) |
423 x 467 x 247 | 567 x 677 x 340 | 694 x 820 x 300 | 1470 x 1260 x 700 |
Head weight | 20 kg | 40 kg | 55 kg | 210 kg |
Electronic unit size | 2U x 19” | |||
Electronic unit weight | 6,5 kg | |||
Max. power consumption | 2500 W | 5000 W | 6000 W | |
Power supply | 115/230 VAC, 1 ph. 50/60 Hz | |||
Remote control | Ethernet, RS-232, IEEE 488 | |||
Operating temperature range (head) |
-10°C to +70°C | +5°C to +70°C |